HOLÁ, Markéta, Pavel KRÁSENSKÝ a Viktor KANICKÝ. Laser Ablation Inductively Coupled Plasma Atomic Emission Spectrometry for Soil Analysis. In International Conference on Inorganic Environmental Analysis - Proceedings of the 4th Conference. Pardubice: Spektroskopická společnost J.M.Marci, Univerzita Pardubice, 2005, s. 88-90.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Laser Ablation Inductively Coupled Plasma Atomic Emission Spectrometry for Soil Analysis
Název česky ICP-OES pro analýzu půd
Autoři HOLÁ, Markéta (203 Česká republika, garant), Pavel KRÁSENSKÝ (203 Česká republika) a Viktor KANICKÝ (203 Česká republika).
Vydání Pardubice, International Conference on Inorganic Environmental Analysis - Proceedings of the 4th Conference, od s. 88-90, 3 s. 2005.
Nakladatel Spektroskopická společnost J.M.Marci, Univerzita Pardubice
Další údaje
Originální jazyk angličtina
Typ výsledku Stať ve sborníku
Obor 10406 Analytical chemistry
Stát vydavatele Česká republika
Utajení není předmětem státního či obchodního tajemství
Kód RIV RIV/00216224:14310/05:00014698
Organizační jednotka Přírodovědecká fakulta
Klíčová slova anglicky Laser ablation; Indctively coupled plasma; Optical emission spectrometry; Soil
Štítky Indctively coupled plasma, laser ablation, Optical emission spectrometry, soil
Změnil Změnila: Mgr. Markéta Holá, Ph.D., učo 4801. Změněno: 3. 2. 2006 16:16.
Anotace
Analysis of soils is very important in environmental and agricultural chemistry. The conventional methods of soil preparation for inductively coupled plasma optical emission spectrometry (ICP-OES) are based on sample dissolution. However, decomposition of some samples is a very complicated and time-consuming procedure. Therefore, ICP spectrometry combined with laser ablation (LA) might be a solution for the analysis of resistant materials such as soils or other geological samples. Soil samples were presented for the analysis as cast pellets with epoxide resin binder. Q-switched Nd:YAG laser operated at the wavelengths of 1064 nm and 266 nm and repetition rate of 10 Hz was used for ablation of prepared pellets. Samples were placed into an ablation cell which was connected with XY-translator allowing a sample moving during ablation. The ablated material was transported by Ar along a polyamide tubing to ICP spectrometer Jobin Yvon 170 Ultrace (laterally viewed). The analytical response for soil pellets were used to produce the signal vs. element content plots using 1064 nm and 266 nm laser wavelenght.
Anotace česky
Byla vypracována metoda pro stanovení vybraných stopových prvků v zemědělských půdách technikou LA-ICP-OES.