RYVOLOVÁ, Markéta, Jan PREISLER, František FORET, Peter HAUSER, Pavel KRÁSENSKÝ, Brett PAULL a Mirek MACKA. Combined Contactless Conductometric, Photometric and Fluorimetric Single Point Detector for Capillary Separation Methods. Analytical Chemistry. Washington, D.C., USA: American Chemical Society, roč. 82, č. 1, s. 129-135. ISSN 0003-2700. 2010.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Combined Contactless Conductometric, Photometric and Fluorimetric Single Point Detector for Capillary Separation Methods
Název česky Kombinovaný bezkontaktní vodivostní, fotometrický a fluorimetrický jednobodový detektor pro kapilární separační metody
Autoři RYVOLOVÁ, Markéta (203 Česká republika, domácí), Jan PREISLER (203 Česká republika, garant, domácí), František FORET (203 Česká republika), Peter HAUSER (756 Švýcarsko), Pavel KRÁSENSKÝ (203 Česká republika, domácí), Brett PAULL (372 Irsko) a Mirek MACKA (372 Irsko).
Vydání Analytical Chemistry, Washington, D.C., USA, American Chemical Society, 2010, 0003-2700.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10406 Analytical chemistry
Stát vydavatele Spojené státy
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 5.874
Kód RIV RIV/00216224:14310/10:00042881
Organizační jednotka Přírodovědecká fakulta
UT WoS 000273265700027
Klíčová slova anglicky detection; capillary separation methods; combined detector; photometric detection; fluorimetric detection; C4D
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Ing. Andrea Mikešková, učo 137293. Změněno: 18. 7. 2013 15:39.
Anotace
This work for the first time combines three on-capillary detection methods, namely capacitively coupled contactless conductometric (C4D), photometric (PD) and fluorimetric (FD), in a single (identical) point of detection cell, allowing concurrent measurements at a single point of detection for use in capillary electrophoresis, capillary electrochromatography and capillary/nano-liquid chromatography. The novel design is based on a standard 6.3 mm i.d. fiber-optic SMA adapter with a drilled opening for the separation capillary to go through, to which two concentrically positioned C4D detection electrodes with a detection gap of 7 mm were added on its each side acting simultaneously as capillary guides. The optical fibers in the SMA adapter were used for the photometric signal (absorbance) and another optical fiber at a 45 degree angle to the capillary was applied to collect the emitted light for FD. Light emitting diodes (255 nm and 470 nm) were used as light sources for the PD and FD detection modes. LOD values were determined under flow-injection conditions to exclude any stacking effects: For the 470 nm LED LODs for FD and PD were for fluorescein (1x10-8 mol/L) and tartrazine (6x10-6 mol/L), respectively and LOD for the C4D was for magnesium chloride (5x10-7 mol/L). The advantage of the three different detection signals in a single point is demonstrated in capillary electrophoresis using model mixtures and samples including a mixture of fluorescent and non fluorescent dyes and common ions, underivatized amino acids, and a fluorescently labeled digest of bovine serum albumin.
Návaznosti
GA203/09/1025, projekt VaVNázev: Multidetekční platforma pro mikrokolonové separace proteinů a peptidů
Investor: Grantová agentura ČR, Multidetekční platforma pro mikrokolonové separace proteinů a peptidů
MSM0021622415, záměrNázev: Molekulární podstata buněčných a tkáňových regulací
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Molekulární podstata buněčných a tkáňových regulací
VytisknoutZobrazeno: 28. 3. 2024 20:25