Silver(I) Ions Ultrasensitive Detection at Carbon Electrodes-Analysis of Waters, Tobacco Cells and Fish Tissues
KŘÍŽKOVÁ, Soňa; Olga KRYŠTOFOVÁ; Jaromír HUBÁLEK; Libuše TRNKOVÁ; Vojtěch ADAM et al.Základní údaje
Originální název
Silver(I) Ions Ultrasensitive Detection at Carbon Electrodes-Analysis of Waters, Tobacco Cells and Fish Tissues
Název česky
Ultrasenzitivní detekce stříbrných iontů na uhlíkových elektrodách - analýza vody, tabákových buněk a rybích tkání
Autoři
KŘÍŽKOVÁ, Soňa; Olga KRYŠTOFOVÁ; Jaromír HUBÁLEK; Libuše TRNKOVÁ; Vojtěch ADAM; Miroslava BEKLOVÁ; Aleš HORNA; Ladislav HAVEL a René KIZEK
Vydání
Sensors, Basel, Switzerland, Molecular Diversity Preservation Interna, 2009, 1424-8220
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10405 Electrochemistry
Stát vydavatele
Švýcarsko
Utajení
není předmětem státního či obchodního tajemství
Impakt faktor
Impact factor: 1.821
Označené pro přenos do RIV
Ano
Kód RIV
RIV/00216224:14310/09:00037454
Organizační jednotka
Přírodovědecká fakulta
UT WoS
Klíčová slova anglicky
silver; guppy (Poecilia reticulata); tobacco cells; ecotoxicology; voltammetry; miniaturized carbon electrodes
Změněno: 8. 12. 2009 11:07, RNDr. Kamila Neplechová, Ph.D.
V originále
We used carbon paste electrodes and a standard potentiostat to detect silver ions. The detection limit (3 Signal/Noise ratio) was estimated as 0.5 mu M. A standard electrochemical instrument microanalysis of silver(I) ions was suggested. As a working electrode a carbon tip (1 mL) or carbon pencil was used. Limits of detection estimated by dilution of a standard were 1 (carbon tip) or 10 nM (carbon pencil). Further we employed flow injection analysis coupled with carbon tip to detect silver(I) ions released in variousbeverages and mineral waters. During first, second and third week the amount of silver(I) ions releasing into water samples was under the detection limit of the technique used for their quantification. At the end of a thirteen weeks long experiment the content of silver(I) ions was several times higher compared to the beginning of release detected in the third week and was on the order of tens of nanomoles. In subsequent experiments the influence of silver(I) ions (0, 5 and 10 mu M) on a plant model system (tobacco BY-2 cells) during a four-day exposition was investigated. Silver(I) ions were highly toxic to the cells, which was revealed by a double staining viability assay. Moreover we investigated the effect of silver(I) ions (0, 0.3, 0.6, 1.2 and 2.5 mu M) on guppies (Poecilia reticulata). Content of Ag(I) increased with increasing time of the treatment and applied concentrations in fish tissues. It can be concluded that a carbon tip or carbon pencil coupled with a miniaturized potentiostat can be used for detection of silver(I) ions in environmental samples and thus represents a small, portable, low cost and easy-to-use instrument for such purposes.
Česky
K detekci stříbrných iontů byly použity uhlíkové elektrody. Detekční limit byl stanoven na 0,5 mikroM. Systém průtokové analýzy spojené s uhlíkovou elektrodou byl použit ke stanovení Ag ve vodě a různých nápojích, dále ve tkáních ryby paví očko a v modelovém rostlinném systému tabákových buněk BY-2.
Návaznosti