Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine

KELBL, Martin; Ján KOČIŠ; Radek VESELÝ; Zdenek FLORIAN; Tomas NAVRAT et al.

Základní údaje

Originální název

Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine

Název česky

Biomechanické testování systému Arcofix na prasečí páteři

Autoři

KELBL, Martin; Ján KOČIŠ; Radek VESELÝ; Zdenek FLORIAN; Tomas NAVRAT a Petr VOSYNEK

Vydání

Asian Spine Journal, Uijeongbu-si, Korean Society of Spine Surgery, 2015, 1976-1902

Další údaje

Jazyk

angličtina

Typ výsledku

Článek v odborném periodiku

Obor

30211 Orthopaedics

Stát vydavatele

Korejská republika

Utajení

není předmětem státního či obchodního tajemství

Odkazy

Označené pro přenos do RIV

Ano

Kód RIV

RIV/00216224:14110/15:00083910

Organizační jednotka

Lékařská fakulta

EID Scopus

Klíčová slova česky

přední stabilizace; biomechanika;stabilizace páteře; zlomeniny hrudní a bederní páteře

Klíčová slova anglicky

Anterior stabilization; Biomechanics; Spine stabilization; Thoracolumbar fracture

Štítky

Příznaky

Mezinárodní význam, Recenzováno
Změněno: 28. 4. 2016 14:10, Ing. Mgr. Věra Pospíšilíková

Anotace

V originále

Study Design: An in vitro biomechanical study. Purpose: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. Overview of Literature: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. Methods: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. Results: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. Conclusions: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.

Česky

Některé typy zlomenin hrudní a bederní páteře vyžadují přední stabilizaci. Dlaha Arcofix představuje poslední generaci úhlově stabilního typu dlahy. V uvedené studii jsme testovali biomechanické vlastnosti této dlahy.