KELBL, Martin, Ján KOČIŠ, Radek VESELÝ, Zdenek FLORIAN, Tomas NAVRAT a Petr VOSYNEK. Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine. Asian Spine Journal. Uijeongbu-si: Korean Society of Spine Surgery, 2015, roč. 9, č. 4, s. 503-510. ISSN 1976-1902. Dostupné z: https://dx.doi.org/10.4184/asj.2015.9.4.503.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine
Název česky Biomechanické testování systému Arcofix na prasečí páteři
Autoři KELBL, Martin (203 Česká republika, garant, domácí), Ján KOČIŠ (703 Slovensko, domácí), Radek VESELÝ (203 Česká republika, domácí), Zdenek FLORIAN (203 Česká republika), Tomas NAVRAT (203 Česká republika) a Petr VOSYNEK (203 Česká republika).
Vydání Asian Spine Journal, Uijeongbu-si, Korean Society of Spine Surgery, 2015, 1976-1902.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30211 Orthopaedics
Stát vydavatele Korejská republika
Utajení není předmětem státního či obchodního tajemství
WWW URL
Kód RIV RIV/00216224:14110/15:00083910
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.4184/asj.2015.9.4.503
UT WoS 000420349000002
Klíčová slova česky přední stabilizace; biomechanika;stabilizace páteře; zlomeniny hrudní a bederní páteře
Klíčová slova anglicky Anterior stabilization; Biomechanics; Spine stabilization; Thoracolumbar fracture
Štítky EL OK
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Ing. Mgr. Věra Pospíšilíková, učo 9005. Změněno: 28. 4. 2016 14:10.
Anotace
Study Design: An in vitro biomechanical study. Purpose: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. Overview of Literature: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. Methods: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. Results: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. Conclusions: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.
Anotace česky
Některé typy zlomenin hrudní a bederní páteře vyžadují přední stabilizaci. Dlaha Arcofix představuje poslední generaci úhlově stabilního typu dlahy. V uvedené studii jsme testovali biomechanické vlastnosti této dlahy.