JADCZYK, Tomasz, Jiří WOLF, Martin PEŠL, Filip SOUČEK, František LEHAR, Jiří JEŽ, Tomáš KULÍK, Bohdan TYSHCHENKO, Silvie BĚLAŠKOVÁ, Petr OUŘEDNÍČEK, Guido CALUORI, Miroslav NOVÁK a Zdeněk STÁREK. Reduced Radiation Exposure Protocol during Computer Tomography of the Left Atrium Prior to Catheter Ablation in Patients with Atrial Fibrillation. Diagnostics. Basel: MDPI, 2022, roč. 12, č. 3, s. 1-12. ISSN 2075-4418. Dostupné z: https://dx.doi.org/10.3390/diagnostics12030612.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Reduced Radiation Exposure Protocol during Computer Tomography of the Left Atrium Prior to Catheter Ablation in Patients with Atrial Fibrillation
Autoři JADCZYK, Tomasz, Jiří WOLF (203 Česká republika), Martin PEŠL (203 Česká republika, domácí), Filip SOUČEK (203 Česká republika), František LEHAR (203 Česká republika), Jiří JEŽ (203 Česká republika), Tomáš KULÍK (203 Česká republika), Bohdan TYSHCHENKO, Silvie BĚLAŠKOVÁ (203 Česká republika, domácí), Petr OUŘEDNÍČEK (203 Česká republika, domácí), Guido CALUORI (380 Itálie, domácí), Miroslav NOVÁK (203 Česká republika) a Zdeněk STÁREK (203 Česká republika, garant).
Vydání Diagnostics, Basel, MDPI, 2022, 2075-4418.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30224 Radiology, nuclear medicine and medical imaging
Stát vydavatele Švýcarsko
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 3.600
Kód RIV RIV/00216224:14110/22:00125484
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.3390/diagnostics12030612
UT WoS 000776880600001
Klíčová slova anglicky computed tomography; catheter ablation; radiation
Štítky 14110119, 14110513, podil, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Tereza Miškechová, učo 341652. Změněno: 27. 3. 2023 13:25.
Anotace
(1) Background: Computer tomography (CT) is an imaging modality used in the pre-planning of radiofrequency catheter ablation (RFA) procedure in patients with cardiac arrhythmias. However, it is associated with a considerable ionizing radiation dose for patients. This study aims to develop and validate low-dose CT scanning protocols of the left atrium (LA) for RFA guidance. (2) Methods: 68 patients scheduled for RFA of atrial fibrillation were sequentially assigned to four groups of ECG-gated scanning protocols, based on the set tube current (TC): Group A (n = 20, TC = 33 mAs), Group B (n = 18, TC = 67 mAs), Group C (n = 10, TC = 135 mAs), and control Group D (n = 20, TC = 600 mAs). We used a 256-row multidetector CT with body weight-dependent tube voltage of 80 kVp (<70 kg), 100 kVp (70–90 kg), and 120 kVp (>90 kg). We evaluated scanning parameters including radiation dose, total scanning procedure time and signal-to-noise ratio (SNR). (3) Results: The average effective radiation dose (ED) was lower in Group A in comparison to Group B, C and D (0.83 (0.76–1.10), 1.55 (1.36–1.67), 2.91 (2.32–2.96) and 9.35 (8.00–10.04) mSv, p < 0.05). The total amount of contrast media was not significantly different between groups. The mean SNR was 6.5 (5.8–7.3), 7.1 (5.7–8.2), 10.8 (10.1–11.3), and 12.2 (9.9–15.7) for Group A, B, C and D, respectively. The comparisons of SNR in group A vs. B and C vs. D were without significant differences. (4) Conclusions: Optimized pre-ablation CT scanning protocols of the LA can reduce an average ED by 88.7%. Three dimensional (3D) models created with the lowest radiation protocol are useful for the integration of electro-anatomic-guided RFA procedures.
Návaznosti
MUNI/A/1462/2021, interní kód MUNázev: Nové zobrazovací, výpočetní a analytické metody v diagnostice a monitoraci kardiovaskulárních onemocnění
Investor: Masarykova univerzita, Nové zobrazovací, výpočetní a analytické metody v diagnostice a monitoraci kardiovaskulárních onemocnění
VytisknoutZobrazeno: 26. 4. 2024 18:44