Calibration graphs for Ti, Ta and Nb in sintered tungsten carbide by infrared laser ablation inductively coupled plasma atomic emission spectrometry
KANICKÝ, Viktor, Vítězslav OTRUBA a Jean-Michel MERMETZákladní údaje
Originální název
Calibration graphs for Ti, Ta and Nb in sintered tungsten carbide by infrared laser ablation inductively coupled plasma atomic emission spectrometry
Autoři
KANICKÝ, Viktor (203 Česká republika, garant), Vítězslav OTRUBA (203 Česká republika) a Jean-Michel MERMET
Vydání
Fresenius Journal of Analytical Chemistry, Berlin, Springer-Verlag, 2001, 0937-0633
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10406 Analytical chemistry
Stát vydavatele
Česká republika
Utajení
není předmětem státního či obchodního tajemství
Impakt faktor
Impact factor: 1.649
Kód RIV
RIV/00216224:14310/01:00004732
Organizační jednotka
Přírodovědecká fakulta
UT WoS
000172742900007
Klíčová slova anglicky
Laser ablation; LINA spark; Infrared laser; Inuctively coupled plasma; Atomic emission spectrometry
Štítky
Změněno: 28. 6. 2007 16:42, prof. RNDr. Viktor Kanický, DrSc.
Anotace
V originále
Infrared laser ablation (IR-LA) was studied as a sample introduction technique for the analysis of sintered cobalt-cemented tungsten carbide materials by inductively coupled plasma atomic emission spectrometry (ICP-AES). A Q-switched nanosecond Nd:YAG laser (LINA-Spark Atomizer, LSA Sarl, Cully, Switzerland) based on the Surelite I-20 laser (Continuum, USA) and a Perkin-Elmer Optima 3000 DV ICP system (lateral viewing) were used. The fractionation of cobalt was observed. Linearity of calibration plots was verified at least up to 15 % Ti, 8 % Ta and 3 % Nb. Above 1% (mass %) Ti, Ta and Nb, the repeatability of results was better than 3 % R.S.D. The relative uncertainty at the centroid of the calibration line was in the range from ą 3% to ą 4% for Ti, Ta and Nb with internal standardization by tungsten and up to ą 5% without internal standardization. The limits of detection were 0.004 % Ti, 0.001 % Ta and 0.004 % Nb. The elimination of the cemented hardmetal dissolution procedure is the main advantage of this method.